MDR Certification Process with a Notified Body (NB)

MDR Certification Process

The certification according to the MDR Regulation (Medical Device Regulation) requires a series of complex phases, that must be followed in collaboration with a Notified Body (NB).

Initial Contacts and Submission of the “Pre-Application”

The process starts with initial contacts between the manufacturer or the EU authorised representative and the Notified Body. During this step, a conformity assessment is requested for the concerned medical device. The NB examines the preliminary data provided in the “pre-application” to confirm the classification of the product according to Annex VII of the MDR. In this phase, the manufacturer shall provide detailed information on the medical device concerned by the certification.

Review of the Pre-Application and Quotation

Subsequently, the NB proceeds with the “pre-application” review. This includes the verification  of all preliminary information to ensure that the device falls into the MDR framework and that its classification is correct. Based on this information, the NB prepares a quotation and requests additional clarifications from the manufacturer, if needed.

Official Application Submission

Once the preliminary phase is completed, the manufacturer submits the official application. In this phase, manufacturers must transmit the required documentation in compliance with Annex IX §2.1 or Annex XI Part A §6.1 for the evaluation of the quality management system (QMS). Some NBs can provide checklists to facilitate the process. For class III devices and certain class IIB devices it is requested to submit the required documentation as per Annex IX Chapter II §4.2.

Contract and Application Review

A phase for contracts and agreements follows the application submission. The NB provides an agreement including the applicable terms and conditions. Once both parties have signed the agreement, the NB carefully assesses the submitted documentation to check that it is complete and complying with the regulatory requirements.

Conformity Assessment

At this point, the NB develops a plan of conformity assessment. The complexity of this phase depends on the classification of the device  and on the chosen assessing procedure. The plan can include a series of activities aimed at verifying the compliance of the device and to ensure that it meets all regulatory requirements.

Final Review and Certificate Release

After the conformity assessment, the final review starts and the NB takes its decision. This phase includes a detailed analysis of all reports, documents, and any non conformity solution. The NB takes into account several factors, such as the outcome of post-market surveillance activities and clinical follow-up plans. If everything is in order, the NB releases the CE certificate, in compliance with the information requested by Annex XII of the MDR.

Surveillance Activities

Finally, once the certificate has been obtained, the manufacturer must fulfil the obligations of post-market surveillance, including the continuous monitoring of the device, the post-market clinical follow-up and the management of any significant change in the device. Continuous vigilance is essential to ensure the safety and efficacy of the medical device in time.